Triggering 9-3
The only difference between front panel operation (Figure 9-1) and remote operation (Figure
9-2) is within the idle state of the instrument. Nomenclature in Figure 9-1 relates to the various
names used for configuration menu items, while Figure 9-2 provides the SCPI commands to
control operation.
Arm-In
Event
Arm Event
Detector
Trigger Event
Detector
Trigger Delay
MEASURE
Action
No
Yes
Idle
Another
Arm
?
Another
Trigger
?
Yes
No
Arm
Layer
Trigger
Layer
= GPIB Default
= Output Trigger
Trigger-In
Source
0.0 sec
No
See Note
INITiate
?
Yes
SOURce
ARM
:DIRection
ACCeptor
ARM:SOURce IMMediate
ARM:SOURce BUS
ARM:SOURce TIMer
ARM:SOURce MANual
ARM:SOURce TLINk
ARM:SOURce NSTest
ARM:SOURce PSTest
ARM:SOURce BSTest
TRIGger
:DIRection
SOURce
ACCeptor
TRIGger:DELay <n>
TRIGger:OUTPut
SENSe | NONE
TRIGger:COUNt <n>
ARM:OUTPut
TRIGger|NONE
ARM:COUNt <n>
TRIGger:SOURce IMMediate
TRIGger:SOURce TLINK
Note:
The following commands place the Model 6514 into
idle: ABORt, *RST, SYSTem:PRESet, *RCL <NRf>,
DCL, and SDC.
TRIGger:DELay:AUTO <b>
1
1
Figure 9-2
Trigger model —
remote operation

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals