EasyManuals Logo
Home>Keithley>Measuring Instruments>6514

Keithley 6514 User Manual

Keithley 6514
334 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
11-8 Digital I/O, Analog Outputs, and External Feedback
Preamp out
The preamp output of Model 6514 follows the signal amplitude applied to the input terminal
The preamp output provides a guard output for volts measurements. It can be used as an invert-
ing output or with external feedback in the amps and coulombs modes. Connections and equiv-
alent circuits for the preamp output are shown in Figure 11-6. Full-range outputs for the various
functions and ranges are listed in Table 11-3.
Model 1683
Test Lead kit
RS232DIGITAL I/O
PREAMP
250V PK
2V
OUTPUT
COMMON CHASSIS
120
FUSE LINE
630mA
LINE RATING
50, 60Hz
60 VA MAX
T
(SB)
100 VAC
120 VAC
315mAT
(SB)
220 VAC
240 VAC
INPUT 250V PK
IEEE-488
(CHANGE IEEE
WITH FRONT PANEL MENU)
TRIGGER LINK
!
!
MADE IN
U.S.A.
V, GUARD
(PROGRAMMABLE)
OFF
ON
GUARD
(FOLLOWS
INPUT)
(INTERNAL)
INPUT PREAMP
10K
PREAMP
OUT
2V
OUTPUT
COM
Fuse
Model 6514 Rear Panel
R
L
= Input Resistance of
measuring device
Input from
Prescaler
COM
S
+
-
S
0.1
10k
Model6514A
Measuring Device
(i.e. Chart recorder)
2V Analog Output
A. Connections
LO
HI
B. Equivalent Circuit
Figure 11-5
Typical 2V
analog output
connections

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals