Guarded connections — Connections for guarded volts and ohms measurements are shown
in Figure 3-3. The driven guard (GRD) must be enabled for these measurements.
WARNING With an open input, up to 250V peak may be present on the guard terminals
while in Volts or Ohms. To prevent this, make sure zero check is enabled
whenever the input is open.
Step 6 Disable zero check and take a reading from the display
V-Drop and I-Source for ohms
Model 6514 performs ohms measurement by sourcing a known test current through the DUT
and then measuring the voltage drop across it. The resistance reading is then calculated
(R = V/I) and displayed.