19-4 Calibration
Table 19-1
Recommended calibration equipment
Description Manufacturer/model Specifications
Calibrator Fluke 5700A DC voltage:
1
2V: 7ppm
20V: 5ppm
200V: 7ppm
DC current:
1
20µA: 550ppm
200µA: 100ppm
2mA: 55ppm
20mA: 55ppm
Resistance calibrator Fluke 5450A Nominal resistance:
1
1.9kΩ: 8ppm
1.9MΩ: 11.5ppm
Calibration standard Keithley Model 5156 Nominal resistance:
2
100MΩ: 200ppm
1GΩ: 300ppm
10GΩ: 400ppm
100GΩ: 500ppm
Nominal capacitance:
2
1nF: 1,000ppm
100nF: 1,000ppm
Triax cable Keithley 7024-3
Low-noise coax cable Keithley 4801
Triax-to-alligator clip cable
3
Keithley 237-ALG-2
Triax-to-BNC adapter Keithley 7078-TRX-BNC
Triax shielding cap Keithley CAP-31
BNC-to-double banana plug
adapter
Pomona 1269
1
90-day, 23˚ ±5˚C full-range accuracy specifications shown.
2
23˚ ±3˚C accuracy of characterization.
3
Short red and black clips to make triax short.