19-20 Calibration
Displaying the calibration count
To display the calibration count at any time:
1. From normal display, press SHIFT then CAL. The unit will display the following:
CAL: RUN
2. Use either RANGE key to select CAL:COUNT from the calibration menu, then press
ENTER. For example:
COUNT: 1
3. Press EXIT to return to normal display.