12-2 Remote Operation
Selecting and configuring an interface
Interfaces
Model 6514 supports two built-in remote interfaces:
GPIB interface
RS-232 interface
You can use only one interface at a time. At the factory, the GPIB bus is selected. You can
select the interface only from the front panel. The interface selection is stored in non-volatile
memory; it does not change when power has been off or after a remote interface reset.
GPIB interfaceThe GPIB is the IEEE-488 interface. Model 6514 must be assigned to
unique address. At the factory the address is set to 14, but can be set to any value from 0 to 30.
However, the address must not conflict with the address assigned to other instruments in the sys-
tem. You can use either the SCPI or DDC language to program the instrument.
RS-232 interface When using the RS-232 interface, you must set baud rate, data bits, par-
ity, terminator and flow control. For the RS-232 interface, you can only use the SCPI language
to program the instrument.
Languages
For the GPIB interface, there are two programming languages to choose from:
SCPI language
DDC language
NOTE For the RS-232 interface, only the SCPI language can be used to program the instru-
ment. When the RS-232 interface is selected, it automatically defaults to SCPI.
SCPI language — Standard Commands for Programmable Instrument (SCPI) is fully
supported by the GPIB and RS-232 interfaces. Always calibrate Model 6514 using the SCPI
language.
DDC language — Model 6514 implements most DDCs (device-dependent commands)
available in the Keithley Models 6512, 617, and 617 HIQ electrometers. The commands are pro-
vided in Appendix D. See the appropriate instruction manual for details on operation.
Interface selection and configuration procedures
When you select (enable) the GPIB interface, the RS-232 interface disables. Conversely,
selecting (enabling) the RS-232 interface disables the GPIB interface.

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals