17-2 SCPI Reference Tables
General notes
Brackets ([ ]) are used to denote optional character sets. These optional characters do not
have to be included in the program message. Do not use brackets in the program
message.
Angle brackets (< >) are used to indicate parameter type. Do not use angle brackets in
the program message.
The Boolean parameter (<b>) is used to enable or disable an instrument operation. 1 or
ON enables the operation, and 0 or OFF disables the operation.
Upper case characters indicated the short-form version for each command word.
Default parameter — Listed parameters are both the *RST and SYSTem:PRESet
defaults, unless noted otherwise. Parameter notes are located at the end of each table.
Ref — Refers you to the section (Sec) that provides operation information for that com-
mand or command subsystem.
SCPI — A checkmark () indicates that the command and its parameters are SCPI con-
firmed. An unmarked command indicates that it is a SCPI command, but does not con-
form to the SCPI standard set of commands. It is not a recognized command by the SCPI
consortium. SCPI confirmed commands that use one or more non-SCPI parameters are
explained by notes.
Table 17-1
CALCulate command summary
Command Description
Default
parameter
Ref SCPI
:CALCulate[1] Path to configure and control CALC1 calculations. Sec 7
:FORMat <name> Select math format; MXB or PERCent. MXB
:FORMat? Query math format.
:KMATh Configure math calculations:
:MMFactor <NRf> Set “m” for mX+b calculation; -9.99999e20 to
9.99999e20.
1.0
:MMFactor? Query “m” factor.
:MBFactor <NRf> Set “b” for mX+b calculation; -9.99999e20 to
9.99999e20.
0.0
:MBFactor? Query “b” factor.
:MUNits <name> Specify units for mX+b result; 3 letters (A - Z). “MXB”
:MUNits? Query units.
:PERCent Configure percent math calculation:
[:REFerence] <NRf> Specify reference value; -9.99999e20 to
9.99999e20.
1.0
:ACQuire Use input signal as reference value.
:PERCent? Query reference value.
:STATe <b> Enable or disable CALC1 calculation. OFF

Table of Contents

Other manuals for Keithley 6514