DISPlay, FORMat, and SYSTem 16-3
C) DISPlay:TEXT:STATe <b>
When the text message mode is enabled, a defined message is displayed. When disabled, the
message is removed from the display.
GPIB operation — A user-defined message remains displayed only as long as the instrument
is in remote. Taking the instrument out of remote (by pressing LOCAL or sending the GTL (go
to local) command) or cycling power cancels the message and disables the text message mode.
RS-232 operation — A user-defined message can be cancelled by sending SYSTem:LOCal,