Calibration 19-11
20pA-2µA range calibration
1. Connect the triax shielding cap to the Model 6514 INPUT jack.
Table 19-4
20µA-20mA range amps calibration summary
Model 6514 range Calibrator currents* Settling time**
20µA 0µA >15 sec.
+20.00000µA >15 sec.
-20.00000µA >15 sec.
200µA 0µA >15 sec.
+200.0000µA >15 sec.
-200.0000µA >15 sec.
2mA 0mA >15 sec.
+2.000000mA >15 sec.
-2.000000mA >15 sec.
20mA 0mA >15 sec.
+20.00000mA >15 sec.
-20.00000mA >15 sec.
* Calibrate zero, positive full-scale, and negative full-scale for each range. Triax cap used for zero
cal points.
** Allow calibration signal to settle for indicated time before calibrating each point.