2-4 Measurement Concepts
Input configurations
As shown in Figure 2-1, the input connector can be configured in two ways. With guard off
(Figure 2-1A), input low is connected to the inner shell of the connector. This configuration is
used for Amps, Coulombs, unguarded Volts and unguarded Ohms measurements.
With guard on (Figure 2-1B), the driven guard is connected to the inner shell of the triax con-
nector. Input low is accessed via the COMMON terminal through an internal 0.1 fuse. This
configuration is used for guarded Volts and guarded Ohms measurements only. The GRD key
toggles guard on and off.
NOTE The state of guard (on or off) has no affect on the Amps and Coulombs functions. The
unguarded configuration is always selected for the Amps and Coulombs functions.
Input High
Input Low
INPUT
250V PEAK
Chassis Ground
Volts, Amps, Ohms & Coulombs
A. Unguarded (GRD off)
Input High
Guard
INPUT
250V PEAK
Chassis
Ground
Volts and Ohms only
COMMON
<1
Input Low
B. Guarded (GRD on)
Fuse
Figure 2-1
Input connector
configurations

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals