EasyManuals Logo
Home>LINET>Medical Equipment>Multicare X

LINET Multicare X User Manual

LINET Multicare X
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #127 background imageLoading...
Page #127 background image
D9U001MCX-0101_03
127
17.7.10 Fault Signalization
The system is protected against failure states, by stopping and braking the drive system, and respective signalization. The fault
indicator ashing briey and the accumulator indicator shows the fault state. Some defects are cleared automatically (e.g.: drive
overheating). When drive or electronics is overheated, an short acoustic signal occurs before the drive is blocked.
ERROR LED1 LED2 LED3 LED4
Drive overheated OFF OFF OFF ON
Electronics overheated OFF OFF ON OFF
Brake error OFF OFF ON ON
Retraction not completed OFF ON OFF OFF
Internal system error OFF ON OFF ON
Closing of the Field-e󰀨ect transistor is penetrated OFF ON ON OFF
Control circuit overheated OFF ON ON ON
Controlcircuiterror ON OFF OFF OFF
Activation button stuck ON OFF OFF ON
Retraction button stuck ON OFF ON OFF
A button on the main control panel is pressed when activating the i-Drive
Power system by ON button
ON OFF ON ON
17.7.11 Light Indicators
Indicator Meaning
Go Indicator
Constantly lit
Flashing
Hand is on touch sensor; i-Drive Power wheel is ready for use.
Hand is not on touch sensor; i-Drive Power is not ready for use.
Fault Indicator
Constantly lit
Flashing
i-Drive Power cannot be activated (i-Drive Power wheel is not lowered, castor control lever is
braked, bed is connected to the mains).
System is faulty (indicated on accumulator status indicator).
OR
i-Drive Power control box heat protection is activated
17.7.12 Technical Specications
Parameter Value
i-Drive Power wheel diameter 21 cm
Max. fast forward speed (at ground, loaded) 4,43 Km/h (±15%)
Max. forward speed (at ground, loaded) 2,16 Km/h (±15%)
Max. reverse speed (at ground, loaded) 2,16 Km/h (±15%)
Max. angle of ascent
Noise level (when retracting the drive wheel) 65 dB