D9U001MCX-0101_03
37
10 Accumulator
WARNING!
When the bed is not connected to the mains and accumulator is not suciently charged all electrical func-
tions of the bed are blocked! After some long period without charging the battery, the battery will be deeply
discharged and loose its full functionality!
Purpose
The accumulator serves as a backup during power failures or for emergency bed positioning.
► Use only accumulator approved by the manufacturer.
► Check the functionality of the accumulator at least once a month and have the accumulator changed if necessary.
The manufacturer will assume no responsibility for any damage to the bed or the accumulator caused by:
■ non-observance of the manufacturer’s instructions in the instructions for use.
■ using accumulator not approved by the manufacturer.
Warranty
The manufacturer provides a 6-month warranty for the full function of the accumulator.
Accumulator lifetime could be up to 5 years if operated under optimum conditions.
Accumulator capacity can be signicantly reduced if:
► too high ambient temperature
► many accumulator charge/discharge cycles
► recurrence of deep discharge
► bed is often powered only by the accumulator
Charging
The accumulator supplied with the bed is delivered insuciently charged.
Charging of the accumulator before use of the bed takes approximately 4 hours.
Charging of the accumulator runs automatically when the power cable is connected to the mains.
To charge the accumulator:
► Connect the bed to the mains.
Storage
For declared lifetime period of leaded accumulators is recommended during storage:
► To prevent accumulator from deep discharging and to keep accumulator at least partly charged by regular recharging.
► To store accumulator on the dry places with temperature from 10°C to 40°C.
► To prevent accumulators from being in the sunshine.
Signalisation
The indicator on Multiboard X touchscreen and LED on Attendant Control Panel indicate the accumulator charge status.