17.7.14 i-Drive Power Maintenance
Periodical maintenance of the i-Drive Power must be done by qualied service technician or authorized service organization at least
once a year. To continue maintenance please see chapter Maintenance.
Service technician must check the following:
accumulator status and eventual replacement of the accumulator (after maximum of three years of duty)
gas spring – replace if necessary (after maximum of three years of duty)
i-Drive Power wheel – replace if necessary
lifting mechanism – grease if necessary
cables, control elements – replace if necessary
i-Drive Power function
17.7.13 Electrical specication
Parameter Value
Input Voltage, Frequency 230 V AC, 50/60 Hz
127 V AC, 50/60 Hz
120 V AC, 50/60 Hz
110 V AC, 50/60 Hz
100 V AC, 50/60 Hz
Accumulator Voltage 36 V DC, Capacity: 9 Ah
Maximum Power Input 300 W
Fuse
Version 230 V
Version 127 V
Version 120 V
Version 110 V
Version 100 V
2 x T1,6A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V
2 x T3,15A L 250V