WARNING!
Do not overload the bed (SWL), respect the duty cycle (INT.) and consider chapter 22 Maintenance in order to
maintain the basic safety with regard to electromagnetic disturbances for the expected service life.
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not o󰀨er
adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as reloca-
ting or re-orienting the equipment.
WARNING!
Use of the accessories, converters and other cables, than specied and provided by manufacturer of this bed 
could lead to increase of electromagnetic emission or lower the electromagnetic immunity of this bed and
lead to improper operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna) should
not be used closer than 30 cm (12 inches) from any part of this bed Multicare X, including cables specied by 
manufacturer. Otherwise this could lead to deterioration of functionality of this bed.
5.6.1 Manufacturer instructions - electromagnetic emissions
Emission Test Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations / icker emissions
IEC 61000-3-3
Complies
List of used cables:
Mains cable, maximum length 6 m
Attendant Control Panel, maximum length 3m
Handset, maximum length 3m
5.6 Electromagnetic Compatibility
Bed is intended for hospitals except for near active HF surgical equipment and the RF shielded room of a medical system
for magnetic resonance imaging, where the intensity of EM disturbances is high.
Bed has dened no essential performance.
WARNING!
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result
in improper operation. If such use is necessary, this equipment and the other equipment should be observed
to verify that they are operating normally.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare X and is the answer not in the manual?

LINET Multicare X Specifications

General IconGeneral
BrandLINET
ModelMulticare X
CategoryMedical Equipment
LanguageEnglish

Related product manuals