D9U001MCX-0101_03
16
VISUAL
SIGNALISATION
LOCK LED
lit
ashing:
0,6s lit / 0,6s not lit not lit
Thighrest, Calfrest and Bed Extension Lock LED locked lock error keyboard error motion blocked unlocked
Backrest Lock LED locked lock error keyboard error motion blocked unlocked
Bed Height, Lateral Tilt, Trendelenburg and Antitrende-
lenburg Tilt Lock LED
locked lock error keyboard error motion blocked unlocked
Foot Control Lock LED locked lock error keyboard error motion blocked unlocked
1.8.5 Lock LED (Attendant Control Panel)
1.9 Denitions
Basic Bed Conguration The pricelist model conguration, not including a mattress.
Bed Weight The value depends on the product conguration, accessories or customer adjustments.
Clearance of Undercarriage The height from the oor to the lowest point of the undercarriage between the castors, for
the manipulation of accessories under a braked bed in the standard position.
Duty Cycle Cycle of operation of the motor: time of activity/time of rest.
Ergoframe Ergoframe is the kinematic system of Mattress support platform Adjustment whose eect is
the elimination of pressure on the patient´s abdomen and pelvic area and frictional forces
on the patient‘s back and legs.
Maximum Patient Weight Maximum Patient Weight depends on the application environment according to IEC 60601-
2-52. For application environment 1 (intensive/critical care) and 2 (acute care) reduce Safe
working Load by 65 kg. For application environment 3 (long-term care) and 5 (ambulatory
care) reduce Safe working Load by 35 kg.
Safe Working Load The highest allowable load on the bed.
The Safe Working Load is the sum of:
– the patient
– the mattress
– the accessories of the medical bed (only if they are supported by the support system of
the medical bed) and
– the load supported by those accessories (excluding patient weight).
Siderail Height The height of the upper crossbar or the edges of the siderails (not the highest point of the
siderail controls) from the patient surface.
Standard Bed Position - The height of the patient surface with regard to the oor is 400 mm.
- The mattress support platform, including the individual parts, has to be in a horizontal
position (0°).
- The siderails are always locked in the upper position.
- The basic position of the integrated extension.
Adult Patient having a physical size equal to or more than 146 cm, a mass equal to or more than
40 kg and a body mass index (BMI) equal to or more than 17 (according to IEC 60601-2-
52).
Maximum Mass of Mobile Hospi-
tal Bed
Sum of Maximum Empty Bed Mass and Safe Working Load.