WARNING!
Risk of injury due to incorrect placement of an infusion pump!
► Ensure the infusion pump on the Foldable infusion stands will not collide with any movable parts of the Multicare
X (especially Backrest) or with the patient!
CAUTION!
Risk of material damage due to incorrect placement of an infusion pump!
► Place an infusion pump carefully on the telescopic part of the Foldable infusion stands in order to prevent the
telescopic part from being damaged!
Foldable infusion stand equipped with 2 hooks is intended for carrying IV bags or baskets for intravenous solutions.
The pair of Foldable infusion stands can serve as handles for bed transport when they are raised.
It is possible to extend height of the Foldable infusion stand and to fold down the Foldable infusion stand again.
Pair of the Foldable infusion stands should be located on the head end Accessory Adaptor.
Maximum load of one hook is 5 kg.
To install Foldable IV & Drive on the head end Accesso-
ry Adaptor:
► Place IV & Drive on the Accessory Adaptor.
► Insert the screw from below so that the clamping
lever points inwards.
► Push the clamping lever upwards.
To remove Foldable IV & Drive from the head end Acces-
sory Adaptor:
► Release the clamping lever by pushing it down.
► Remove the screw from below.
► Remove the IV & Drive from the Accessory Adaptor.
► Place the IV & Drive in a safe position outside the
bed.
19.21 Foldable IV & Drive (Infusion Stands/Pushing Handles)
Fig. Fixation of the Foldable IV & Drive on the Accessory Adaptor
Fig. Position of the Foldable IV & Drive on the Accessory Adaptor