D9U001MCX-0101_03
99
Fig. First degree of the bed overload (displayed Hi)
25. 12. 25
22:58
CURRENT WEIGHT
Max 250 kg, Min 10 kg, e = 0,5 kg
0,5 0,1
Hi
ZERO
T
07
14.5 Bed Overload
If load of the bed is over 254,5 kg:
Hi“ is shown on the display.
If load of the bed is over 260 kg:
Warning popu-up window is displayed over the Scales screen.
NOTE If the bed is overloaded, it is impossible to position or manipulate the bed until overloading is removed. Bed overloading
always has higher priority than Hold Mode and Taring.
HOLD PROCESS INTERRUPTED
Start the process again
Cancelling the HOLD Mode Signalisation & Settings
To cancel the HOLD Mode:
Press Cross Icon on the corresponding HOLD Mode
Guide window.
See the window displayed in the case of the interrupted
HOLD Mode setting on the right side.
HOLD MODE
Attach / Remove all required accessories.