D9U001MCX-0101_03
14
1.7 Acoustic signalisation
SOUND MEANING
CONTINUOUS SOUND overheating
accumulator overcurrent
scales overload (only version with scales)
actuator overload
BEEP + CONTINUOUS SOUND Siderail Signal (lateral tilt + head siderail or foot siderail down)
REPEATED BEEP: 0,6s sound / 2,6s silence STOP error (all STOP buttons are disabled)
MELODY: 3 beeps, pause, 2 beeps, longer pause, 3 beeps,
pause, 2 beeps
Bed Exit Alarm (only version with scales)
BEEP lasting 0,3s conrmation
stopping or locked function
lateral tilt 15° achieved
transition from tilt (lateral tilt, Trendelenburg, Antitrendelenburg)
to horizontal position
4 TIMES REPEATED BEEP lasting 0,3s disconnected from the mains
positioning powered by the accumulator
BEEP lasting 0,5s start of service mode or end of service mode
keyboard error (positioning blocked)
BEEP lasting 3s system error
BEEP lasting 5s SCU disconnected (only if integrated mattress is used)
scale module disconnected (only version with scales)
REPEATED BEEP during 3 minutes: 1,1s sound / 1,1s silen-
ce
Brake Signal (only version with Brake Signal)
MAINS POWER LED MEANING
lit connected to the mains
ashing: 0,6s lit / 0,6s not lit keyboard error (ashing inverted to Lock LED)
error (rst fault)
ashing: 0,1s lit / 0,1s not lit service mode
not lit disconnected from the mains power
transformer switching error
1.8 Visual signalisation
1.8.1 Mains Power LED (Attendant Control Panel)