D9U001MCX-0101_03
155
WARNING!
Risk of injury due to incorrect use!
Multicare X bed is compatible with Traction Frame Kinetec T41 only if the bed is equipped with the Adaptor for
Traction Frames!
There must be no patient on the bed when installing the Traction Frame!
Avoid collisions between the Traction Frame and the bed (Foot Board, Backrest and Lateral Tilt) during bed posi-
tioning!
Avoid collisions between the Traction Frame and accessories!
Cross the thresholds with caution during transporting the bed with installed Traction Frame!
It is allowed to transport a patient on the bed with Traction Frame just in emergency cases and with caution!
Respect the Safe Working Load of the bed, of the Traction Frame and of its hooks and pulleys!
Remove the Traction Frame from the bed if it is not needed for treatment!
WARNING!
Risk of injury due to incorrect use!
Multicare X bed is compatible with Traction Frame M only if the bed is equipped with the Adaptor for Traction
Frames!
There must be no patient on the bed when installing the Traction Frame!
Avoid collisions between the Traction Frame and the bed (Foot Board, Backrest and Lateral Tilt) during bed posi-
tioning!
Avoid collisions between the Traction Frame and accessories!
Cross the thresholds with caution during transporting the bed with installed Traction Frame!
It is allowed to transport a patient on the bed with Traction Frame just in emergency cases and with caution!
Respect the Safe Working Load of the bed, of the Traction Frame and of its hooks and pulleys!
Remove the Traction Frame from the bed if it is not needed for treatment!
19.16 Traction Frame M
19.17 Traction Frame Kinetec T41
Fig. Safe Working Load of the hooks
(Infusion Holder)
Fig. Safe Working Load of the pulley
WARNING!
Read Instructions for use of the Traction Frame Kinetec T41 before use of this product!
WARNING!
Read Instructions for use of the Traction Frame M before use of this product!
Intended Use
Traction Frame is a supporting construction intended for xation, traction and relief of limbs, spine and pelvis.
Traction Frame is intended for orthopaedic department, surgery department, traumatologic department and for ICU.
Placement
Traction Frame is inserted to the holes in the Accessory Adaptor at head end and to the holes in the Adaptor for Traction Frames at
foot end.
Intended Use
Traction Frame is a supporting construction intended for xation, traction and relief of limbs, spine and pelvis.
Traction Frame is intended for orthopaedic department, surgery department, traumatologic department and for ICU.
Placement
Traction Frame is inserted to the holes in the Accessory Adaptor at head end and to the holes in the Adaptor for Traction Frames at
foot end.