EasyManuals Logo
Home>LINET>Medical Equipment>Multicare X

LINET Multicare X User Manual

LINET Multicare X
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
D9U001MCX-0101_03
28
6 Use and Storage Conditions
DANGER!
Danger to life due to electric shock!
To ensure the bed’s class I protection against electric shocks:
Ground the mains.
Use exclusively Hospital Grade or Hospital Only receptacles for grounding.
Multicare X and OptiCare X are designed for use in rooms for medical purposes. Electrical installations must therefore meet local
norms laying down the necessary conditions for electrical installations.
Disconnect the bed from the mains in exceptional cases (i.e. lightnings, earthquake).
Multicare X and OptiCare X are not suitable for indoor environments containing ammable gases (except oxygen cylin-
ders). Respect values of the parameters connected with environment conditions in the chapter Technical Specication 
during use and storage of the product.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare X and is the answer not in the manual?

LINET Multicare X Specifications

General IconGeneral
BrandLINET
ModelMulticare X
CategoryMedical Equipment
LanguageEnglish

Related product manuals