Chapter 3 Horizontal System Setting RIGOL
MSO4000/DS4000 User’s Guide 3-9
Roll Mode
In this mode, the waveform scrolls from the right to the left. The adjustable range of
the horizontal time base is from 200.0 ms/div to 1.000 ks/div. The horizontal time
base will be set to 200.0 ms/div automatically when you turn on the Rollmode.
Note:
In this mode, the To Adjust the Horizontal Position”,To Turn Delayed
Sweep On/Off”,Trigger, To Specify the Measurement Range,
Protocol Decoding”,Pass/Fail Test”,Waveform RecordingandTo
Set the Persistence Time” are not available.
If the delayed sweep function is enabled, when theRoll” mode is enabled, the
delayed sweep function will be disabled automatically; when the “Y-T” mode is
enabled again, the delayed sweep function will be enabled automatically.
Note: Please pay attention to distinguish the Roll mode from “Slow Sweep Mode.