S2/S2BW
Digital Color Doppler Ultrasound System
10.3.3 Adjust PW Sample Volume
When in pre-active or Active PW mode, move the
TrackBall
up and down to adjust the depth of Sample
Volume;
Press
SET
then drag the
TrackBall
to adjust the size of the sample volume; then press
SET
again to confirm
your desired size
10.3.4 Adjust PW Sample Angle
When in pre-active or active PW mode, turn the
ANGLE
key to adjust the correction angle of the sample
volume.
10.3.5 Adjust Baseline
When in pre-active or active PW mode, or after frozen image, turn the
BaseLine
swtich to adjust
Baseline position; turn the switch up to move up baseline, turn down the switch to move down
baseline; notice when baseline is not properly placed, the spectrum graph might look like aliasing.
10.3.6 Dual Display
Dual Display in PW mode allows you to compare different images taken at different time.
Operate:
1.
When in live PW-mode, press
Dual
key to start dual display, now the screen shows two images, on the
left is a PW-mode live image;
2.
Press
L/R
key, to freeze the PW-mode image on the left and switch to a active PW-mode image on the
right;
3.
Press
L/R
again, to switch between the left image andthe right image, when the current image is
active, the transducer marker is displayed in green;
4. When finished, press
Dual
agai to quit Dual Display.
A dual display image is in the following format:
P/N: 4710.00149A01
10-5

Table of Contents

Other manuals for Sonoscape S2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sonoscape S2 and is the answer not in the manual?

Sonoscape S2 Specifications

General IconGeneral
BrandSonoscape
ModelS2
CategoryMedical Equipment
LanguageEnglish

Related product manuals