Page #57 background imageLoading...
Page #57 background image
57
Function Settings
Chapter 5 System Administrator Settings
[Add 0 to Patient ID]
Select whether to add zeros to patient IDs with lengths
shorter than the specified number of digits in the
setting screen that appears, and press the ENTER
button.
The default setting is [Use].
[Patient Name Field]
Select the number of display fields for the patient
name in the setting screen that appears, and press the
ENTER button.
If you select [2] or [3], the patient name field will be
separated into first, middle, and last names.
The default setting is [1].
[Change Patient Info.]
Select whether to allow changes to patient information
at any time in the setting screen that appears, and press
the ENTER button.
The default setting is [Not Used].
[Rec List Display Items]
Select whether the [Rec List Display Items] screen
displays titles or patient information in the setting
screen that appears, and press the ENTER button.
The default setting is [Title Name]. If you select
patient information, patient IDs and patient names will
be displayed.
[Recording Preparation]
Select whether to use the [Recording Preparation]
screen in the setting screen that appears, and press the
ENTER button.
The default setting is [Use].
If you select [Not Used], [Recording Preparation] will
be grayed out and disabled in the [MENU] screen.
3
To configure settings in other tabs, use the V, v, B,
and
b buttons to select a tab.
If you are finished configuring [Function Settings],
use the
V, v, B, and b buttons to select [Apply], and
press the ENTER button to save the configurations.
[General 2] Tab
Configure initial system settings.
1
Use the B and b buttons to select the [General 2] tab.
2
Configure each setting.
Select an item using the
V
and
v
buttons, and press the
ENTER button to display the setting screen for that item.
[Prohibit Button Operation]
Select whether to prohibit operation of the unit via the
buttons on the front panel in the setting screen that
appears, and press the ENTER button.
The default setting is [Not Used].
The prohibit button operation function does not apply
to the MENU,
V, v, B, b, BACK, and ENTER buttons
and the 1 (on/standby) switch.
[CUSTOM Button Control]
Select the function to assign to the CUSTOM button
on the front panel in the setting screen that appears,
and press the ENTER button.
The default setting is [Recording List].
The functions that can be assigned are as follows.
Tip
Assignable
function
Description
Not Used A function is not assigned.
Freeze ON Pause the video at any position
during playback.

Table of Contents

Other manuals for Sony HVO-500MD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sony HVO-500MD and is the answer not in the manual?

Sony HVO-500MD Specifications

General IconGeneral
DisplayLCD
Product colorWhite
LED indicatorsYes
Playback resolutions supported1280 x 720
Storage temperature (T-T)-20 - 60 °C
Operating temperature (T-T)5 - 40 °C
Operating relative humidity (H-H)20 - 80 %
HDD capacity500 GB
BNC input ports1
USB 2.0 ports quantity4
Optical drive includedNo
Weight and Dimensions IconWeight and Dimensions
Weight2900 g
Dimensions (WxDxH)212 x 290 x 105.5 mm