To use
Use the Ampl tab to select automatic amplitude measurements.
1. Select a source with the Channels Selector.
2. Click a Measurements button to add an Amplitude measurement to the Measure list:
Measurement buttons
Amplitude Low +Ovrshoot
AC RMS Max -Ovrshoot
Pk-Pk Min Mean
RMS Cycle RMS Cycle Mean
High
3. Click Display On in the upper-right corner of the Measurement Setup Control Window to turn the measurement display on
and off.
Behavior
Measurements are updated approximately three times a second, or less for long record lengths. Measurements are calculated
after each acquisition. You can display up to eight automatic measurements.
TIP. Select the individual measurements directly from the Measure menu Amplitude submenu.
What do you want to do next?
Learn about taking timing measurements.
Learn about taking histogram measurements.
Measurement setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 273

Table of Contents

Other manuals for Tektronix MSO73304DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO73304DX and is the answer not in the manual?

Tektronix MSO73304DX Specifications

General IconGeneral
BrandTektronix
ModelMSO73304DX
CategoryTest Equipment
LanguageEnglish

Related product manuals