Set up an SPI serial bus
First, follow the common steps to set up any bus. To continue to set up an SPI serial bus, follow these steps:
1. To define the bus type in the Bus Setup window, click Serial as the Bus Type, and select SPI from the drop-down list.
2. To select the input for the displayed components of the bus, click on a component Input and select a channel from the
cascading list.
3. To set an input voltage reference threshold level, double-click the Threshold entry box and use the keypad to enter the level.
4. To set the input polarity, click the Polarity entry box, and select Rising or Falling or Active Low or Active High from the drop-
down lists.
5. To set the input word size, double-click the Word Size entry box, and use the keypad or multipurpose knobs to enter the
size.
6. To set the input direction, click the Bit Order entry box, and select MS First or LS First from the drop-down list.
7. Click the Display tab, and configure the bus display.
Set up an I2C serial bus
First, follow the common steps to set up any bus. To continue to set up an I2C serial bus, follow these steps:
1. To define the bus type in the Bus Setup window, click Serial as the Bus Type, and select I2C from the drop-down list.
2. To select the input for the displayed components of the bus, click on the component Input and select a channel from the
cascading list.
3. To set an input voltage reference threshold level, double-click the Threshold entry box and use the keypad to enter the level.
4. Click the Display tab, and configure the bus display.
Set up an RS-232 serial bus
NOTE. This online help file supports many oscilloscope models from Tektronix. This feature is only available on some models.
First, follow the common steps to set up any bus. To continue to set up an RS-232 serial bus, follow these steps:
1. To define the bus type in the Bus Setup window, click Serial as the Bus Type, and select RS-232 from the drop-down list.
2. To select the input for the displayed components of the bus, click on a component Input and select a channel from the
cascading list.
3. To set an input voltage reference threshold level, double-click the Threshold entry box and use the keypad to enter the level.
4. To set the input polarity, click the Polarity entry box, and select Normal or Inverted from the drop-down lists.
5. To set the input data bits, double-click the Data Bits drop-down list, and select use the keypad or multipurpose knobs to
enter the size.
6. To set the parity, click Parity entry box, and select None, Odd, or Even from the drop-down list.
7. To set the bit rate, click Bit Rate entry box, and select Custom, 50, 300, 1200, 2400, 9600, 19200, 38400, 115200, 921600,
or 10000000 from the drop-down list.
8. Click the Display tab, and configure the bus display.
Set up a USB serial bus
First, follow the common steps to set up any bus. To continue to set up an USB serial bus, follow these steps:
1. To define the bus type in the Bus Setup window, click Serial as the Bus Type, and select USB from the drop-down list.
2. To select the input for the displayed components of the bus, click on a component Input and select a channel from the
cascading list.
How to ?
582 DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series

Table of Contents

Other manuals for Tektronix MSO73304DX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO73304DX and is the answer not in the manual?

Tektronix MSO73304DX Specifications

General IconGeneral
BrandTektronix
ModelMSO73304DX
CategoryTest Equipment
LanguageEnglish

Related product manuals