Assembly
6 Requirements
The room chosen for installation must satisfy the
following requirements:
Closed, dry room
Clean, level and sufficiently stable subsurface
There should be no large interference fields
present (e.g. strong magnetic fields) that could
interfere with correct operation of the unit.
The required ambient conditions are satisfied
(refer to the "Technical Data").
7 Installation
7.1 Establishing the electrical
connections
The connection ports are located in the recess
on the rear of the unit.
Requirements:
ü
Correctly installed power outlet in the vicinity of
the unit (max. length of mains cable 3 m)
ü
The plug connection of the power supply is
freely accessible so that it can be quickly dis-
connected in the event of danger.
ü
Mains voltage must match the information
shown on the type plate of the power supply
unit
ü
The supply voltage of the power supply unit
matches the data on the type plate.
Plug in the connecting plug of the connecting
cable into the connecting socket of the unit.
Fit the correct country-specific adapter.
Plug the mains plug into the power outlet.
Assembly
9000-615-28L02 1903V004 17
EN

Table of Contents

Other manuals for Durr Dental Vector Paro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental Vector Paro and is the answer not in the manual?

Durr Dental Vector Paro Specifications

General IconGeneral
BrandDurr Dental
ModelVector Paro
CategoryDental equipment
LanguageEnglish

Related product manuals