EasyManuals Logo

Durr Dental Vector Paro User Manual

Durr Dental Vector Paro
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
15.3 Changing the valve in the
fluid container
The valve on the underside of the fluid container
must be cleaned regularly and checked for
blockages and leaks.
Press the valve lightly against the fluid con-
tainer and unscrew it anti-clockwise.
Clean the valve.
If cleaning is not possible, e.g. if the filter in the
valve is blocked, then the valve needs to be
replaced.
Check the O-ring.
In the event of leaks, loose seating or visible
damage the O-ring must be replaced.
Insert the valve in the holder and tighten it
clockwise as far as it will go.
15.4 Checking instrument wear
Instrument wear can be checked using the
Vector toolcard:
Place the handpiece against the toolcard. If
the instrument tip extends beyond the red
marking then the instrument can still be used.
If the instrument tip just reaches the red mark-
ing then the instrument shows signs of wear
but can still be used.
If the instrument tip does not reach the red
marking then the instrument must be replaced.
Usage
9000-615-28L02 1903V004 59
EN

Table of Contents

Other manuals for Durr Dental Vector Paro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental Vector Paro and is the answer not in the manual?

Durr Dental Vector Paro Specifications

General IconGeneral
BrandDurr Dental
ModelVector Paro
CategoryDental equipment
LanguageEnglish

Related product manuals