13.3 Preparation at the operating
location
Wear protective gloves.
Wear protective goggles.
Use a mask.
Use protective clothing.
WARNING
Risk of infection from contaminated
products
Danger of cross contamination
❯
Reprocess the product correctly and
promptly before its first use and after
every subsequent use.
Rinsing the handpiece with water
Rinsing with water will rinse any remaining Vector
polish residue from the handpiece and prevent
blockages.
❯
Fill the fluid container to approx. 1/3 with water.
❯
Place the handpiece in a sink or in a suitable
container.
❯
Start the rinsing process:
Press the RINSE button for at least 2 seconds.
❯
The LED flashes – the rinsing process takes
around 30 seconds and ends automatically.
Pre-cleaning
Do pre-cleaning of handpiece and acces-
sories no more than 15 minutes after the
unit has been used.
❯
Clean the exterior surfaces completely with two
cleaning cloths . Make sure that the sur-
faces are sufficiently moistened.
❯
Note the action time of the cleaning agent.
❯
Perform the procedure twice.
❯
Draw 3 x 20 ml cold water (temperature
< 20 °C) into a conventional sterile 20-ml dis-
posable pipette with Luer connection and rinse
the inner lumen of the handpiece.
Transport
❯
Protect the device from contamination when
you transport it from the treatment chair to the
reprocessing location.