11 Treatment
11.1 Preparation
Perform the following steps before every treat-
ment:
❯
Make sure that only handpieces and instru-
ments are used that have been reprocessed
since the last treatment.
❯
Check the nozzle for correct operation,
"Checking the operation of the nozzle".
❯
Check that the sealing ring on the instrument
chuck and the support seal on the union nut
are correctly seated and in perfect condition,
"Instrument chuck/sealing ring check".
❯
Check that the instrument is correctly seated
and in perfect condition, "Inserting/changing
instruments".
❯
Check the fill level of the fluid container.
❯
Adjust the power on the operating panel as
required, "POWER".
11.2 Treatment using the Paro
handpiece
During treatment using the Paro handpiece fluid
is applied as a pulsating jet. Once the flexible foot
switch is released a small amount of fluid will still
drip out.
CAUTION
Risk of burns if no fluid comes out
During operation the resonating body
and the instrument move. If fluid does
not come out, friction heat can build up
under direct contact between the res-
onating body or the instrument and dry
mucous membranes, which can cause
burns.
❯
Only operate the handpiece with the
nozzle inserted and working correctly.
❯
Only operate the handpiece with the
rotary adaptor and ring cover in proper
working order.
❯
Activate the handpiece by operating the flexible
foot switch.