14 Treatment breaks for more
than 24 hours
If no treatment is carried out for a period of
24 hours or more, reprocessing of the hose sys-
tem must be performed.
14.1 Cleaning and disinfecting the
hose system
The hose system is disinfected using the ready-
to-use, aldehyde-free solution Vector/RinsEndo
Disinfectant.
Preparation:
Before starting the CLEAN or RINSE (rins-
ing/disinfecting) functions, if there is a
fluid bag in the unit it will need to be
removed. If an attempt is made to start
the function with a fluid bag inserted, a
long, low audible signal will sound and the
action will not be performed.
❯
Remove the cover of the fluid bag.
❯
Push the piercing mechanism to the rear.
❯
Take out the fluid bag.
❯
Slide the piercing mechanism forwards as far
as it will go.
❯
Pull off the hose connection from the hand-
piece.
❯
Lay the handpiece hose in the sink.
Rinsing with water:
❯
Fill the fluid container to approx. 1/3 with
water.
❯
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds.
❯
LED flashes – the rinsing process takes
around 30 seconds and ends automatically.
❯
Rinsing with water will flush out any remain-
ing Vector Polish residue.
❯
After the end of the rinsing process drain any
remaining fluid in the system.
Disinfection with Vector/RinsEndo Disinfec-
tant:
❯
Pour 2 sealing caps (approx. 40 ml) of the
Vector/RinsEndo Disinfectant in the fluid
container.
Usage
56 9000-615-28L02 1903V004
EN