Performing pairing between the flexible foot
switch and the device:
Switch on the device .
3 s
1.
2.
LIQUID
PEDAL
FUNCTION
CLEANRINSE
POWER
LIQUID
PEDAL
FUNCTION
LED PEDAL flashes in orange.
Press the flexible foot switch for around 3
seconds until the orange PEDAL LED goes
out.
Result:
After successful pairing the unit is immediately
ready for operation.
If nothing happens when the
flexible foot
switch is operated, it is possible that the
foot switch being used is paired with a
different device. In this case cancel the
pairing and repeat the process.
Cancelling a pairing
Requirements:
ü
No foot switch cable must be connected to
the device or to the flexible foot switch.
ü
The orange PEDAL LED is not on or is not
flashing.
Switch off the device
.
Touch the LIQUID operating panel, keep touch-
ing it and switch on the device .
Result:
When the orange PEDAL LED flashes, this shows
that the current pairing has now been cancelled.
8 Commissioning
8.1 Function check
To finish the initial start-up process, all of the
connections must be checked to make sure they
are securely seated and leak-tight.
Check for correct operation:
Operating panel
Flexible foot switch
Visual and acoustic signals
Assembly
9000-615-28L02 1903V004 19
EN

Table of Contents

Other manuals for Durr Dental Vector Paro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental Vector Paro and is the answer not in the manual?

Durr Dental Vector Paro Specifications

General IconGeneral
BrandDurr Dental
ModelVector Paro
CategoryDental equipment
LanguageEnglish

Related product manuals