Areas that are difficult to reach, e.g. the
Paro handpiece instrument chuck, should be
thoroughly rinsed (at least 5 x for 5 seconds
each) using an air and water syringe.
Drying
❯
If necessary, re-dry at a clean location using a
hygienic, lint-free cloth.
❯
Blow dry the components with compressed air
in a clean location.
13.7 Automatic cleaning, interme-
diate rinsing, disinfection,
final rinse, drying
Selection of the washer-disinfector
Automatic cleaning and disinfection requires a
washer-disinfector with the following properties
and validated processes:
– Corresponds to and tested in accordance with
EN ISO 15883
– Certified program for thermal disinfection (A
0
value ³ 3000 or at least 5 minutes at 93°C)
– Programme is suitable for the components
and provides sufficient rinsing cycles.
For more information: "General information".
Selection of the machine cleaning agents and
disinfectants
The following properties are required:
– Material compatibility with the product
– Corresponds with the manufacturer's specifi-
cations of the CD
For further information, see: "General informa-
tion".
Automatic cleaning and disinfecting
When arranging the parts in the washer-
disinfector, make sure there are no areas
missed by rinsing.
❯
Place the handpiece on the special mountings
for transmission instruments (e. g. Miele:
ADS 2, Æ approx. 16 mm) in the washer-disin-
fector.
❯
Place scaler instruments onto the special
mountings for instruments (e.g. Miele: A 814) in
the washer-disinfector.
❯
Place Paro instruments in the instrument tray
and place in the basket for small parts.
❯
Fix removable parts of the handpiece (ring
cover, rotary adaptor, scaler cover, light con-
ductor), instrument holders of the tool kits
(without instruments) and torque wrench with a
suitable washer-disinfector holder.
13.8 Check for function
❯
After the end of the cleaning and disinfection
cycle, check the components for any residual
soiling and moisture. If necessary, repeat the
cycle.
❯
Check the components for damage and
replace if necessary.
❯
The parts should be packaged as soon as
possible after drying and checking.