Areas that are difficult to reach, e.g. the
Paro handpiece instrument chuck, should be
thoroughly rinsed (at least 5 x for 5 seconds
each) using an air and water syringe.
Drying
If necessary, re-dry at a clean location using a
hygienic, lint-free cloth.
Blow dry the components with compressed air
in a clean location.
13.7 Automatic cleaning, interme-
diate rinsing, disinfection,
final rinse, drying
Selection of the washer-disinfector
Automatic cleaning and disinfection requires a
washer-disinfector with the following properties
and validated processes:
Corresponds to and tested in accordance with
EN ISO 15883
Certified program for thermal disinfection (A
0
value ³ 3000 or at least 5 minutes at 93°C)
Programme is suitable for the components
and provides sufficient rinsing cycles.
For more information: "General information".
Selection of the machine cleaning agents and
disinfectants
The following properties are required:
Material compatibility with the product
Corresponds with the manufacturer's specifi-
cations of the CD
For further information, see: "General informa-
tion".
Automatic cleaning and disinfecting
When arranging the parts in the washer-
disinfector, make sure there are no areas
missed by rinsing.
Place the handpiece on the special mountings
for transmission instruments (e. g. Miele:
ADS 2, Æ approx. 16 mm) in the washer-disin-
fector.
Place scaler instruments onto the special
mountings for instruments (e.g. Miele: A 814) in
the washer-disinfector.
Place Paro instruments in the instrument tray
and place in the basket for small parts.
Fix removable parts of the handpiece (ring
cover, rotary adaptor, scaler cover, light con-
ductor), instrument holders of the tool kits
(without instruments) and torque wrench with a
suitable washer-disinfector holder.
13.8 Check for function
After the end of the cleaning and disinfection
cycle, check the components for any residual
soiling and moisture. If necessary, repeat the
cycle.
Check the components for damage and
replace if necessary.
The parts should be packaged as soon as
possible after drying and checking.
Usage
9000-615-28L02 1903V004 53
EN

Table of Contents

Other manuals for Durr Dental Vector Paro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental Vector Paro and is the answer not in the manual?

Durr Dental Vector Paro Specifications

General IconGeneral
BrandDurr Dental
ModelVector Paro
CategoryDental equipment
LanguageEnglish

Related product manuals