Preferably insert the "Paro straight probe" into
the chuck and secure using the tool kit cover.
Switch on the device
.
Start the handpiece with the foot switch.
During treatment using the Paro handpiece the
nozzle ejects fluid as a pulsating stream. The
fluid hits the instrument in the top third.
Possible faults:
If spray mist is released there is possibly air
in the hose lines.
Nozzle blocked; replace the nozzle.
Stream of fluid does not meet the instrument
in the top third or completely misses the
instrument; replace the nozzle.
Instrument chuck/sealing ring check
During treatment, two sealing rings prevent fluid
from entering the instrument chuck.
Support seal in the union nut
Sealing ring at the instrument chuck
Usage
22 9000-615-28L02 1903V004