– Scaler instrument P4, 120 mm, for the
removal of supragingival coatings on smooth
surfaces and for interdental areas.
POWER
2–3 LEDs: 40 % - 60 % operating perfor-
mance
11.5 After every treatment
End of treatment
❯
Dismantle the instrument with the tool kit cover.
❯
After every treatment, all parts used must be
cleaned, disinfected and, if necessary, ster-
ilised, "13 Reprocessing".
12 Cleaning
12.1 Cleaning of the outside sur-
faces
All outside surfaces must be cleaned and disin-
fected if they are contaminated or soiled.
– Surface of the device
– Handpiece hose
– Protective cap of the fluid polish bag
– Fluid container
– Tool kit for Scaler without instruments
As the surface disinfectant, we recommend using
a disinfectant that is compatible with the materi-
als and meets the general dental hygiene stan-
dards, e.g.:
– Dürr Dental FD 322 rapid surface disinfectant
– Dürr Dental FD 350 disinfection wipes
– Dürr Dental FD 366 sensitive rapid surface dis-
infectant
NOTICE
Liquid can cause damage to the unit.
❯
Do not spray the unit with cleaning and
disinfectant agents.
❯
Make sure that liquid does not get
inside the unit.
❯
For pre-cleaning, remove coarse organic soiling
with a paper towel.
❯
Clean the surfaces with a moist, soft, lint-free
cloth.
Usage
40 9000-615-28L02 1903V004