Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds.
❯
LED flashes – the Vector/RinsEndo Disinfec-
tant is rinsed into the system and the pro-
cess ends automatically. The Vector/
RinsEndo Disinfectant stays in the system
until the next treatment.
❯
Empty any remaining Vector/RinsEndo Dis-
infectant from the fluid container.
❯
Thoroughly rinse the fluid container with
water and dry.
14.2 Initial start-up after a break in
treatment for more than 24
hours
The initial start-up process will depend upon
whether the hose system was reprocessed
before the break in treatment. Proceed as follows
depending on the situation:
1. Reprocessing performed before the treat-
ment break:
❯
Rinse the system with water.
Thorough rinsing with water will
flush out
any disinfectant remaining in the hoses,
and any irritation to the patient caused by
the taste of residual disinfectant will be
avoided.
❯
Pull off the hose connection from the hand-
piece.
❯
Lay the handpiece hose in the sink.
❯
Fill the fluid container with water.
❯
Touch the RINSE (rinsing/disinfecting) button
for at least 2 seconds.
LED flashes – the rinsing process takes
around 30 seconds and ends automatically.
Preparing the device for treatment:
❯
Fill the fluid container.
❯
Insert a fluid bag.
❯
Fluid pre-delivery to the handpiece:
Touch the FLUID button for at least 2 sec-
onds.
The LED flashes simultaneously with the LED
in the piercing mechanism; the process ends
automatically.
2. Reprocessing not performed before the
treatment break:
❯
Perform reprocessing before initial start-up of
the unit, "13 Reprocessing".
Usage
9000-615-28L02 1903V004 57