9.6 Service kit
The service kit has been configured for the
Vector System.
The service kit contains replacement parts and
tools that can be used for maintenance and
repairs, "15 Maintenance".
9.7 Fluid container
POWER
RINSE
CLEAN
FLUID
LIQUID
PEDAL
FUNCTION
The fluid container can hold up to approx. 600ml
of water or watery fluids.
❯
The fluid container should be emptied and
cleaned every evening and descaled as
required, "12.3 Cleaning the fluid container".
The
fluid container is not suitable for ther-
mal disinfection or sterilisation.
Water quality
The water quality must meet the general require-
ments for water supply systems in a dental
surgery and all applicable standards.
Active ingredient solutions
NOTICE
Blockage of the nozzle when mixing
water-based active ingredient solu-
tions with Vector Fluid Polish.
This combination leads to flocculation of
the fluid and to blockages as a result.
❯
Vector Fluid Polish must not be added
when using water-based active ingre-
dient solutions.
❯
After using a water-based active ingre-
dient the unit should be thoroughly
rinsed using warm water.
In order to widen the range of possible treat-
ments, additional active ingredients can be
added to the water.
Possible active ingredients include e.g. chlorhexi-
dine digluconate solutions up to a total concen-
tration of 0.2 %; other solutions may damage the
system and can lead to unsuccessful treatment.
9.8 Vector Fluid Polish
Usage