15.11 Changing the sealing ring of
the resonant ring
Sealing ring, order number
NOTICE
Damage to the resonant ring due to
twisting
When loosening and tightening the nut
do not hold down the resonant ring.
Dismantle the handpiece, "Disassembly".
Place the sealing ring and SW 5 flat wrench
from the service kit ready.
~45°
Loosen the nut using the flat wrench, turning
approx. 45° (1/4 turn) anti-clockwise.
Change the sealing ring of the resonant ring.
2
1
~45°
3
Tighten the nut using the flat wrench, turning
approx. 45° (1/4 turn) clockwise.
Mount the handpiece, "Assembly".
Usage
66 9000-615-28L02 1903V004