Set-up the machine
Connect/Confirm Concentrates
CAUTION
The operator must make sure that the chosen concentrates on the
Information Display correspond with the connected concentrates
during function check and during treatment.
Verify that the prescribed concentrate(s) for the specific treatment is
(are) used. Avoid changing type of concentrate containers during
treatments.
CAUTION
Note
• In order to finalize the function check, the A-concentrate/acetate
must be connected and the venous drip chamber filled to the
correct level. The reason is, that the conductivity cells must
be checked w ith conductivity and the air detector must be
checked with fluid in the chamber. When the function check is
finished, the treatment time appears on the time display.
• During the conductivity test in function check it is not possible
to change between Bicarbonate and Acetate mode. If the
conductivity check during function check was performed in
Acetate mode it is not possible to switch to Bicarbonate mode
during treatment. However, it is always possible to change
from Bicarbonate to Acetate mode during treatment.
• Fixed conductivity alarm limits (±5 %) will be set automatically,
centred around the conductivity set value.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:7