8. Heparin P ump (option)
The Heparin pump can be programmed for different syringe
sizes
. Syringes must comply with ISO 7886-2.
Instructions on how to attach the syringe to the pump, and
how to do the heparin pump settings, can be read in "Arterial
Blood Line - Attach" starting on page 4:16 in part 1, see
point 8.
9. Priming Detector
The priming detector detects if there is blood in the venous
blood line. When blood has been detected, treatment alarms
are activated. T herefore, it is of the utmost importance that
the venous blood line is correctly placed in the priming
detector before treatment is started, see "Venous Blood Line
- Attach" s tarting on page 4:28 in part 1, see point 5 for
instructions on this.
Before the priming detector detects blood certain alarms are
suppressed to facilitate the priming procedure. When blood
has been detected, the treatment time starts to count down
and accumulated treatment parameter values will start to be
measured and displayed. The treatment time will continue to
count down as long as blood is detected.
2:10
AK 96
®
Operator's Manual - Machine Description
HCEN12239 Revision 12.2010
Program version 3.xx