Diascan RESCHEDULED TWICE
Press SELECT key to confirm
Appears:
Appears when two Diascan measurement checks in a row
has been interrupted by any of the following actions:
- the blood flow rate is changed
-theBlood pump button is pressed
-theStart UF Stop button is pressed
-theFluid bypass button is pressed
- a treatment alarm which produces dialysis fluid
bypass or blood pump stop is generated
- sodium and bicarbonate (conductivity) of the dialysis
fluid is changed
- temperature of the dialysis fluid is changed
-dialysisfluid flow rate is changed
- UF volume is changed
- linear UF profiling settings is changed
- linear UF profiling activation/deactivation is made
- sodium or bicarbonate profiling activation/deactiva-
tion is made
- isolated UF activation is made
The machine was not able to completely perform the
measurement check as the measurement check has been
interrupted.
Machine actions:
None.
1
4:44
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order