RECOVERED FROM POWER FAILURE
Press SELECT key to confirm
Appears:
When the m achine has recovered from power failure.
Machine actions:
None.
Measures:
Press the Select key. The Time Display shows “reC”
until the Select key has been pressed. See Recovery from
Machine Shut Down" on page 3:29 in part 1 for further
information.
Mute tim e:
2min.
Buzzer sound:
Soft buzzer
1
4:106
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx