BLOOD DETECTED IN ASSISTED PRIM
Appears:
When the priming detector detects blood during assisted
priming procedure.
Machine actions:
The blood pump stops.
Measures:
First, make sure that the patient has not been connected to
the blood lines. If this is not the case, clean the priming
detector lens. A lternatively call the authorized technician.
Mute time:
2min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:15
Attention alarms in alphabetical order