Start the Machine
Check Before Switching the Machine On
That the mains cable is connected to a mains supply with protective
earth.
That the main switch on the rear of the machine is in the on position
(indicated by a lit segment shown on the Time display).
That the water supply is connected to the machine (inlet water
tube) and switched on.
That the drain tube (outlet tube) is properly placed with an air gap
between the machine and the drain/sewer system.
That the dialysis uid tubes are connected to the safety couplings
of the machine.
Switch the Machine On
Note
If the machine has been stored with chemical disinfectant, this
solution must be rinsed out of the system before any treatment
can be initiated.
Note
Switch the machine on by pressing the On/Off button for three seconds.
At this moment, the buzzer will sound shortly and the computer
processor of the machine will count up (displayed on the Time display).
The current program version ins
talled on the machine w ill be displayed
on the Information Display.
After a short while all the lamps on the operator's panel will light up
simultaneously. The operator must look and check that the lamps
are functionin
g. If any lamp does not light up, contact an authorized
service technician. The machine will also test the buzzer sound.
4:2
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx

Table of Contents