13. Stand-by Port for Blue Concentrate Connector
The blue concentrate connector with white tube marking
(point 7 in this list) must be placed in this port at all times
except when used for intake of liquid bicarbonate or acetate
concentrate, during function check and treatment.
The connector is to be attached to the port in the same way
as the yellow disinfectant connector, see point 6 in this list.
14. BiCart® Cartridge Holder (option)
This holder is used for the B iCart cartridge, a cartridge
containing dry bicarbonate powder, for treatments. It is also
used for the CleanCart A or C, cartridges containing cleaning
(A)ordecalcification (C), agents, for heat disinfection
programs with CleanCart cartridge.
Instructions on how to attach the BiCart cartridge to the
holder can be read in "Connect/Confirm Concentrates", page
4:7 in part 1, and how to change it during treatment on page
3:32 in part 1.
Instructions on how to attach the CleanCart cartridge to
the holder can be read in "Heat Disinfection Program with
CleanCart® cartridge - Performing", page 8:13 in part 1.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Machine Description