Single Measuring
The pressure in the blood pressure measurement
cuff will be maximum 180 mmHg
the first time
the blood pressure is being measured when a blood
pressure measurement check is carried out. If there are
difficulties obtaining the correct values there will be two
remeasurement checks with higher cuff pressure. If the
measurement checks still fail there will be an attention
alarm to guide the operator to take correct actions. If
the measurement check was successfully performed, the
cuff pressure will be maximum 30 mmHg above the
latest measured systolic blood pressure for following
measurement checks.
Procedure
1. Press the BPM button.
The BPM MENU will be shown on
the Information Display.
2. When START has been selected,
press the Select key to start the
measurement check.
The BPM MENU will remain on
the Information Display during the
ongoing measurement check. The
current pressure in the blood pressure
measurement cuff is displayed in
P CUFF.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - BPM
11:7