Discontinuing
Note
• If the treatment is to be continued, fi rst press the
flashing Time button to confirm treatment end. The
TREATMENT TIME EXPIRED attention alarm will then be
cleared and the treatment time confirm menu will be displayed
on the Information Display. Then press the Back key and
increase the treatment time (SET)usingtheKeypad.Increase
the UF volume, or set the min UF rate to zero.
• If the treatment is to be discontinued before the set
treatment time is achieved, the remaining time should be
decreased to 0.00 (press Time button, change SET using the
Keypad).
• If the treatment time is decreased to 0:00 (i.e. the set time is
decreased in SET)andthereisstill time left in isolated UF
phase, an attention alarm will appear. Deactivate isolated UF,
press the Select key for 3 seconds when the isolated UF menu
is displayed, or set ISOL UF time to zero.
Note
Confirm Treatment End
When the remaining treatment time shown on the Time Display equals
0:00 an attention alarm appears. Simultaneously, the Time button starts
to flash.
At this moment the UF rate will be automatically reduced to minimum
UF rate as the set UF volume is achieved; the UF Start/Stop button will
be extinguished to indicat
ethis. TheTMPalarmlimitswill
automatically be widened. The time shown on the Time Display will
continue to count down but will show negative figures.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:59