On/Off button
Press and hold the On/Off button for 3 seconds to switch the machine
on and off.
At that moment, when the machine is switched on, the button lights up
and a welcome display is shown on the Information Display for a few
seconds. This display includes information on which program version
is currently installed in the machine.
The On/Off button continues to be lit at all times when the machine
is switched on.
It is not possible to switch the machine off, using the On/Off button,
during treatment when blood is detected.
Note
• The main switch on the rear of the machine should always be
in On position even when the machine is not in use. This is in
order for the batteries to be charged. See "Rear Component
Details"starting on page 2:31 in part 1, see point 13 for further
information.
Note
Priming button
The Priming button automatically lights up when the machine has
been switched on during the function check. T
his indicates that the
machine is ready for blood line attachment and priming of the blood
compartment of the dialyzer.
The Priming button continues t o be lit, indicating that priming mode
is active, until blood is detected.
Press the lit Priming button and the priming main menu opens on the
Information Display.
3:10
AK 96
®
Operator's Manual - Operating the Machine
HCEN12239 Revision 12.2010
Program version 3.xx