Attention Alarm List
AIR DETECTOR NOT ACTIVATED
Press AIR DETECTOR button
Appears:
During priming 15 seconds after the venous drip chamber
has been filled upon attention alarm request.
Machine actions:
The Air detector button will be flashing.
Measures:
Press the flashing Air detector button to activate the air
detector alarm function.
Mute time:
2min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:5
Attention alarms in alphabetical order