Heparin
From the Blood Path main menu; select HEPARIN by using the
Keypad to reach the sub m enu.
The following will be shown on the Information Display:
Type of syringe (e.g. Terumo 30 mL).
BOLUS; set heparin solution bolus volume.
FLOW RATE; set heparin solution ow rate
STOP LIMIT; set stop time in minutes at which the heparin pump
should stop before treatment end
ACC HEP; sum of accumulated bolus volume during priming and total
accumulated volume since treatment start
The heparin pump will run concurrent to the blood pump in double
needle treatment. In single needle treatment it will be running
continuously irrespective of if the blood pump is running or not.
Single needle
From the Blood Path main menu; select SINGLE NEEDLE by using
the Keypad to reach the sub menu.
The following will be shown on the Information Display:
MIN STROKE; displays the low alarm limit for the stroke volume
PV LOW; low venous pressure alarm limit
P VEN; current venous pressure
PV HIGH; high venous pressure alarm limit
SET/MEAN QB; displays set and mean blood ow rates. The
set blood ow rate (arterial) is to be set/adjusted using the
Blood Pump Up/Down keys. The mean blood ow rate is the effective
blood ow rate displayed when performing single needle treatment
3:14
AK 96
®
Operator's Manual - Operating the Machine
HCEN12239 Revision 12.2010
Program version 3.xx