BLOOD LEAK DETECTOR FAILURE
Clean and press SELECT key
Appears:
During function check if the blood leak detector is dirty
or out of function.
Machine actions:
None.
Measures:
Clean the blood leak detector, see "Blood Leak Detector"
on page 8:41 in part 1 for instructions. Press the
Select key when cleaned.
If the attention alarm appears again, call an authorized
technician.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms